Industrial pressure gauge, vacuum gauge and pressure vacuum gauge instrument calibration or instrument calibration procedures - Database & Sql Blog Articles

Shunuo varistor / ESD varistor / product complete / Sunlord first-class agent
Photocoupler

Industrial pressure gauge, vacuum gauge and pressure vacuum gauge instrument calibration or instrument calibration procedures

This procedure is applicable to industrial single-circle spring-loaded pressure gauges, vacuum gauges and pressure vacuum gauges (hereinafter referred to as pressure gauges) after use and repair.

Instrument calibration

Or instrument calibration.
First, the pressure gauge instrument calibration instructions
1. The absolute value of the basic error of the standard shall not exceed 1/3 of the absolute value of the basic error of the pressure gauge being inspected.
2. The test pressure gauge can be carried out at room temperature (5--30 ° C).
3. A pressure gauge with an upper limit of less than or equal to 2.5 kgf/cm2 is used to measure air using air or other inert gas.
4. The accuracy level and allowable basic error of the pressure gauge are as follows:

Pressure gauge calibration requirements 1


5. Working medium is shown in Table 2:

Pressure gauge calibration requirements 2



Second, the instrument appearance inspection
6. The pressure gauge shall be in the case according to the medium to be tested. The liner shall be coated on the dial with the color specified in the following table, and the name of the special medium shall be indicated. The oxygen pressure gauge must be marked with the red “oil-free”.

Pressure gauge calibration requirements 3


7. The components of the pressure gauge shall be assembled firmly and shall not be loosened, and the well shall not have defects such as rust, cracks and holes that affect the measurement performance.
8. Pressure gauge for measuring gas. There should be a safety hole on the back of the case. A dustproof device is required on the safety hole. When the elastic component is broken, the gas can escape from the back of the case.
9. The dial scale numbers and symbols of the pressure gauge should be complete and clear.
The central angle of the re-drawn dial indexing ruler is generally equal to approximately 270°, and the width of the indexing line marked with numbers should not exceed the requirements of Table 4.

Pressure gauge calibration requirements 4



10. The pointer of the pressure gauge should extend into all the indexing lines. The width of the pointer tip should not exceed 1/5 of the minimum indexing interval. The distance between the pointer and the plane of the indexing plate should be in the range of 1--3 mm. The outer diameter of the case is 20 mm or more (including 200 mm). The distance between the pointer and the plane of the indexing plate should be in the range of 2 - 4 mm.
11. The pressure gauge shall be sealed and sealed, and the internal parts shall not be touched without damaging the seal.

Third, the indication verification
12. Zero check: The pressure gauge is in the working position. When there is no pressure and no air, it is carried out before the pressure boosting test and after the pressure reduction test. The following requirements should be met:
(1) The pressure gauge pointer with the zero limit nail should be placed close to the nail, and its "retracted" should not exceed the absolute value of the allowable basic error;
(2) Pressure gauge without zero limit nail, the tip of the pointer shall deviate from the zero index line and shall not exceed the absolute implant allowing the basic error. The zero-point dividing line width is equal to the pressure gauge that allows the absolute value of the basic error, and the pointer must be within the zero-point dividing line width;
(3) The pointer of the pressure vacuum gauge shall be within the width of the zero-point dividing line, and the width of the zero-point dividing line shall not exceed 1.5 times the absolute value of the allowable basic error.
13. The indication test of the pressure gauge is carried out according to the index line marked with numbers. The pressure is gradually increased (or depleted) after the inspection. When the indication value reaches the upper limit of the measurement, it must withstand pressure for 3 minutes (the tube is re-welded) The pressure gauge should be at the upper limit of the measurement for 10 minutes), and then back to the original inspection point in reverse order. The basic error, round-trip difference and tapping displacement should meet the technical requirements of Table 5:

Pressure gauge calibration requirements 5


14. The pressure gauge indicates that the stability of the rotation of the gauge should be smooth and free of jumping or jamming within the full indexing range.
15. The upper limit of the pressure vacuum gauge is below 2.5 kgf/cm2 (including 2.5 kgf/cm2, and the vacuum part is checked by the numbered index line except 760 mmHg; 2.5 kgf/sq. Above the centimeter, the pointer must point to the vacuum part during the evacuation check.

Fourth, additional calibration of several pressure gauges
16. Grease-free check of oxygen pressure gauge:
In order to ensure safety, the oxygen pressure gauge should be certified to be free of grease. Inspection method: Inject pure warm water into the spring tube, shake it, and then pour the water into the appliance containing the living water. There is no colored oil shadow on the water surface, and it can be considered that there is no grease. If the oxygen meter with grease is found to be boiled in boiling water or washed with carbon tetrachloride, confirm that there is no grease and then check.
17. Pressure gauge verification with inspection pointer:
First, the test pointer and the indication finger are simultaneously checked, and the reading is recorded; then the indicator is returned to the zero point, and the indication value is checked again, and the difference between the two boost readings (ie, the check pointer pair) The control force of the indicator shall not exceed the absolute value of the allowable error. The pointer shall not move when tapping the meter.
18. Double-needle double-tube pressure gauges should be checked before the indications are checked. After the two pipes are inspected, the pressure gauge is checked by the three-way joint. A double-needle double- or double-needle single-tube pressure gauge shall also check the difference between the two pointer values, the value of which does not exceed the absolute value of the allowable basic error. The two hands should not affect each other. The two joints should be painted with the same color paint of the same color for identification.
19. Verification of the pressure contact of the charged contact signal device:
The insulation resistance between the telecommunications part of the pressure gauge and the outer casing shall not be less than 20 megohms (500V) when the relative humidity is not more than 80%. The verification method is as follows:
Install the pressure gauge on the checker, use the needle to respectively dial the two signal contact pointers to the outside of the upper and lower limits, and then perform the indication verification. After the indication is passed, the upper and lower signal contact pointers can be set to three. At more than one different verification point, slowly increase or decrease the pressure until the moment when the signal is sent. At this time, the deviation between the reading of the standard pressure gauge and the indication value of the signal pointer shall not exceed the absolute value of the allowable basic error.
Fives,

Instrument calibration

or

Instrument calibration

Result processing
20. Pressure gauges that have passed the instrument calibration or the instrument calibration shall be sealed.
21. The verification period of the pressure gauge shall be determined according to the specific use conditions (generally not more than half a year).
22. If the pressure gauge is unqualified, it can be downgraded for use, but the mark of accuracy level must be changed.

Apple Airtag Screen Protector

This scratch-resistant TPU Film protects your new AirTag from daily wear and tear. It can be easily glued to the back to keep the shiny silver in its original condition. In addition, the AirTag screen protector with "self-healing" function can automatically repair small scratches and bubbles within 24 hours. This means you can continuously enjoy the new look and appearance of AirTag without worrying about serious scratches and dents.
The Apple Airtag TPU Film can perfectly fit your AirTag skin. It is designed to provide the maximum coverage for your AirTag metal panels. Compatible with most AirTag shells. Leave extra space around the frame so that your shell can wrap the edges of the AirTag without interfering with the film.
The 0.14mm ultra-thin AirTag Screen Protector allows you to enjoy the maximum resolution while being thin enough to make the AirTag sensitive to the original signal response, ensuring that it will not affect any user experience. You won't even notice that the screen protector is there, so your Apple logo can still be displayed.
The AirTag Tracker Screen Protector is very easy to apply and is specially designed for your AirTag. Just clean your AirTag and install it on the surface to eliminate all bubbles and get the job done!

Apple Airtag Screen Protector, Apple Airtag Screen Protective Film, Apple Airtag TPU Film, AirTag Tracker Screen Protector, Skin Wrap

Shenzhen Jianjiantong Technology Co., Ltd. , https://www.jjttpucuttingplotter.com

Posted on